国精品无码专区一区二区三区_亚洲av无码在线播放_蜜臀精品无码国产AV一区二区_国产呦在线观看欧美一区

PIOM with expansion screen

Our Product > PIOM > PIOM with expansion screen

PIOM with expansion screen

PIOM-XP
Professional Intraoperative Monitoring System
Channels: 16/32
Functions: SEP/MEP/VEP/BAEP/EEG/EMG

Technique specifications 

AMPLIFIER

1 Number of channels: amplifier 16/ 32 channels.

2 supports up to 32 channels of signal acquisition, with multiple channels integrated into a single output port for connection to consumables, and supports quick connection of acquisition electrodes.

3 Noise level: ≤ 2.0 μV p-p (0.2-200Hz)

4 Input impedance: ≥ 3000M

5 With filtering function: high frequency filtering, 10HZ-7000HZ, low frequency filtering 0.1-500Hz, multi-position adjustable.

6 Common mode rejection ratio ≥ 120dB.prevent accidental vocal cord injury, and the electrode placement ring is designed for improved muscle contact


Software Features

? Computer operating system software: Windows 10 Pro system, professional version of Office software. 

? Nerve monitoring software items: EEG, EMG, cortical localization, somatosensory evoked potentials(SEP), motor evoked potentials(MEP), automatic vertebral pedicle stimulation program, near nerve detection function, auditory evoked potentials(AEP), visual evoked potentials(VEP), etc.

? Multiple inspections are carried out simultaneously on the same screen and can be switched freely.

? Various monitoring modes for all modules can be edited and added according to the needs of the operation, and can be saved as a template to facilitate future monitoring of similar operations.

? TOF test function

? Noise analysis shielding software

? With multiple data windows display: Real-time waveforms, trend graphs, data tables, video images, event windows, etc., it can be displayed on the same screen, or can be viewed window by window in separate screens.

? Report: Template function, can generate report with one click, user can edit and save it by themselves.

? Screen printing function: The monitoring waveform can be copied to the screen and automatically imported into the report or saved as a picture forma